Technical Specifications
PARAMETER | SPECIFICATION |
Lead | Standard 12 leads |
Lead acquisition | 12bit/1000Hz(synchronously 12 leads) |
Work mode | MAN, AUTO, ANALYSIS Modes |
Filter | AC Filter : 50Hz / 60Hz
EMG Filter : 25Hz / 45Hz Anti-Drift Filter 😕 0.15Hz (Adaptive) |
CMMR | ?100dB (with AC FILTER) |
Input circuit | Floating; Protection circuit against
Defibrillator effect |
Input Impedance | ?50M? |
Input? circuit current | ?0.05?A |
Patient current leakage | ?10?A |
Calibrating voltage | 1mV?2? |
Voltage tolerance | ??500mV |
Time constant | ?3.2s |
Frequency response | 0.05?160Hz(-3db) |
Noise level | ?15?Vp-p |
Inter-channel
Interference |
?0.5mm |
Sensitivity | 5?10?20, 40, Auto (mm/mV) |
Recording mode | Default Man. Mode is 3 ch format
Default Auto. Mode is 3 ch+ format |
Paper speed | 6.25?12.5?25?50mm/s??3?? |
Paper | 80mm *20m Roll paper |
Communication | RS232/ USB port |
Power Supply | AC? 110-230V(??10%),50/60Hz????1Hz),60VA
DC Built-in rechargeable battery,14.4V(2000mmAh) |
Dimension/Weight | 300mm?230mm?72mm/2.8Kg |
Reviews